Augmented Marker Tracking for Peri-acetabular Osteotomy Surgery: A Cadaver Study

4 pagesPublished: June 13, 2017

Abstract

Purpose. To validate a small, easy to use and cost-effective augmented marker-based hybrid navigation system for peri-acetabular osteotomy surgery.
Methods. A cadaver study including 3 pelvises (6 hip joints) undergoing navigated PAO was performed. Inclination and anteversion of two navigation systems for PAO were compared during acetabular reorientation. The hybrid system consists of a tracking unit which is placed on the patient’s pelvis and an augmented marker which is attached to the patient’s acetabular fragment. The tracking unit sends a video stream of the augmented marker to the host computer. Simultaneously, the augmented marker sends orientation output from an integrated inertial measurement unit (IMU) to the host computer. The host computer then computes the pose of the augmented marker and uses it (if visible) to compute acetabular orientation. If the marker is not visible, the output from the IMU is used to update the orientation. The second system served as ground truth and is a previously developed and validated optical tracking-based navigation system.
Results. Mean absolute difference for inclination and anteversion (N = 360) was 1.34 degrees and 1.21 degrees, respectively. The measurements from our system show a very strong correlation to the ground-truth optical tracking-based navigation system for both inclination and anteversion (0.9809 / 0.9711).
Conclusion. In this work we successfully demonstrated the feasibility of our system to measure inclination and anteversion during acetabular reorientation.

Keyphrases: computer assisted surgery, navigation system, orthopedic surgery

In: Klaus Radermacher and Ferdinando Rodriguez Y Baena (editors). CAOS 2017. 17th Annual Meeting of the International Society for Computer Assisted Orthopaedic Surgery, vol 1, pages 54-57.

BibTeX entry
@inproceedings{CAOS2017:Augmented_Marker_Tracking_Peri,
  author    = {Silvio Pflugi and Till Lerch and Rakesh Vasireddy and Nane Boemke and Moritz Tannast and Timo Michael Ecker and Klaus Siebenrock and Guoyan Zheng},
  title     = {Augmented Marker Tracking for Peri-acetabular Osteotomy Surgery: A Cadaver Study},
  booktitle = {CAOS 2017. 17th Annual Meeting of the International Society for Computer Assisted Orthopaedic Surgery},
  editor    = {Klaus Radermacher and Ferdinando Rodriguez Y Baena},
  series    = {EPiC Series in Health Sciences},
  volume    = {1},
  publisher = {EasyChair},
  bibsource = {EasyChair, https://easychair.org},
  issn      = {2398-5305},
  url       = {/publications/paper/zn},
  doi       = {10.29007/9mbb},
  pages     = {54-57},
  year      = {2017}}